User guide
Help
Every screen, control, color and number in the app, in the order you meet them. Operators can stop after Account. Admins and installers read on. Screenshots and videos come from the demo fleet (MM-1001 to MM-1012). Click any screenshot to enlarge it.
Start here
Magnet Monitor watches MRI magnets and the rooms around them. A small collector at each site reads the magnet's monitor (the MagMon), and optionally room sensors and a UPS, once a minute. The app shows the fleet, opens alerts when a reading goes wrong, and emails or pushes them to whoever is on call.
The five-minute tour
- Open the Fleet page. Units that need attention are sorted to the top. Red is an alarm, amber a warning, green nominal, gray no data, blue maintenance.
- Read the pills on a card. Each pill names one fault and its value, for example Helium critical 22.9 %.
- Click the card. The asset page shows the open alerts, live readings, device error codes, the helium refill forecast and 24 hours of trends.
- Open Alerts. Anything under Needs someone has no owner. Engineers and admins acknowledge or mute from here.
- Check the Debrief each morning. It lists what opened, cleared and stayed open since 9:00 AM yesterday.
Who can do what
Each person has one role per company. Two extra grants, TV and Docs, are ticked per person by an admin.
| Screen or action | Viewer | Engineer | Admin |
|---|---|---|---|
| Fleet, asset pages, Alerts, Debrief, Account | Yes | Yes | Yes |
| Acknowledge, mute, flag false alarm | No | Yes (mute up to 2 weeks) | Yes (any length) |
| TV wall | With TV grant | With TV grant | Yes |
| Docs (installer runbook) | With Docs grant | With Docs grant | Yes |
| Admin panel and Handbook | No | No | Yes |
A superadmin is an admin in every company, and also sees the Companies tab and the company switcher.
Signing in


Sign in with an existing account
- Type your Username and PIN.
- Leave Keep me signed in on this device ticked on your own device: you stay signed in for 30 days. Untick it on a shared computer: the session ends when the browser closes, or after one day.
- Press Sign in.
Create an account with an invite code
- Press Need an account? Register with an invite code.
- Choose a username and a PIN, and type the invite code your admin gave you.
- Press Create account. You are signed in as a Viewer. An admin can raise your role afterward.
PIN rules
- At least 6 characters, at most 72. Letters are allowed.
- Not one character repeated (
111111). - Not a simple run up or down (
123456,987654,abcdef). - Not a common choice such as
123123,passwordorqwerty.
An older, weaker PIN still signs you in, but every page shows a banner asking you to change it under Account.
Wrong PIN and lockout
Any failed sign-in shows the same message: Invalid username or PIN. After several wrong attempts, sign-in from this network pauses for 15 minutes. Five wrong PINs within 15 minutes pause that account from your network for 15 minutes. Wait, or ask an admin to reset the PIN.
View live demo at the top right opens the simulated fleet without signing in.
Fleet dashboard

The status line
Under the title: a green dot and Live, the time of the last update, and the count of assets and sites. The page refreshes itself every 30 seconds while the tab is visible. If it reads Not updating with an amber dot, the last successful refresh was over 75 seconds ago and the readings on screen may be old.
The four tiles

The tiles count connectivity only, not faults.
- Online: a new reading arrived within the unit's offline limit (30 minutes unless an admin changed it).
- Offline: the last new reading is older than that limit.
- Unknown: the unit has never sent a reading.

The alert ticker
The red strip scrolls every unit in alarm or warning with its faults. Hover to pause it. Click a name to open that unit. It disappears when the fleet is clean.
Order of the cards
Alarm first, then warning, nominal, no data, and maintenance last. Within a group the order is alphabetical. There is no manual sort or search.
Reading a card




| Part of the card | What it tells you |
|---|---|
| Name and site | The asset tag and where it is. The small weather icon and temperature are the outside conditions at the site (US National Weather Service). Information only. |
| Colored rail and ring | The alarm level: Alarm Warning Nominal No data Maintenance. The ring's dot pulses while the unit is online. |
| Pills | Up to three faults, worst first. The asset page shows all of them. |
| Six tiles | Helium level (%), Water flow (gpm), Helium pressure (psi), Water temp (°F), Shield temp (K), Compressor (Running, Stopped or Signal lost, with its code). Each has a one-hour sparkline. A dash means no reading. |
| Footer chips | Connectivity (Online, Offline, Unknown), any collector or network chip, and the Cryo chip when the cryogenics are not nominal. |
| Bottom right | Age of the last new reading, or never reported. |
Fault pills and their limits
These are the built-in limits. They are checked only while the unit is online. An admin's alert rules can move the warning lines for a company or a single unit.
| Pill | Shown when | Level |
|---|---|---|
| Compressor stopped code N | CS1 above 0 | Alarm |
| Compressor signal lost code N | CS1 is 20, 30, 31, 32 or 33 and coldhead still at or below 6 K | Warning |
| Coldhead hot | above 20 K | Alarm |
| Coldhead high | above 6 K | Warning |
| Helium critical | below 30 % | Alarm |
| Helium low | below 50 % | Warning |
| Helium high | above 90 % | Warning |
| He pressure high | above 4.5 psi | Alarm |
| He pressure abnormal | below 0 or above 3 psi | Warning |
| Shield hot | above 80 K | Alarm |
| Shield warm | above 60 K | Warning |
| Offline age | No new reading for longer than the offline limit | Alarm |
| Reporting stalled age | The collector's Pi is reachable, but no new data is arriving | Alarm |
Other pills come from open alerts: sensor blind (no reading from a water sensor), Cooling fault, trending (days to a limit), Helium falling (% per day), flatlined (a stuck value), a channel alarm from a rule, and POWER OUTAGE on UPS battery, which is always an alarm.
Footer chips


| Chip | Meaning |
|---|---|
| Online / Offline / Unknown | Connectivity, as counted by the tiles. |
| MagMon silent | The unit is online, but magnet readings have stopped arriving. |
| MagMon not connected | The unit has never sent magnet readings. |
| Env silent | Room sensor and UPS readings have stopped. |
| iR305 / iR305 down | State of the site's cellular router. Shown only where the app can see the router. |
| Tailscale / Tailscale down | State of the collector Pi's VPN link. |
| Cryo watch / urgent / critical / no data | The cryogenics grade. No chip means nominal. |
The router and VPN chips explain why a unit is offline. They never open an alert of their own.
The Cryo grade
One judgment of the cold system, for MRI units only.
| Reading | Watch | Urgent | Critical |
|---|---|---|---|
| Helium level | below 60 % | below 50 % | below 30 % |
| Coldhead | 5.0 K or more | 6.0 K or more | 20 K or more |
| Vessel pressure | 2.0 psi or more | above 3.0 psi | above 4.5 psi |
| Coldhead minus recondenser | - | 2.0 K or more | - |
| Two of coldhead, recondenser, shield at 100 K or more | - | - | Magnet is warm |
The refill chip
The droplet chip estimates when helium will reach 30 %. It uses a straight-line fit of the last 7 days, restarted after any refill. It appears only when helium is clearly falling.
- refill due: under one day.
- 5d to refill: under 14 days. The chip is amber.
- 3w to refill, 3mo to refill: further out. The chip is blue.
- No chip: stable, rising, or 120 days or more away.
Rooms and power (units with sensors or a UPS)
Units fitted with room sensors add one column per zone, Engineering, Tech / Patient and Equipment, each with temperature (°F) and humidity (%RH). Units with a UPS add a power row: Wall power (green), ON BATTERY (red) or Power unknown (gray: the UPS is not answering, which is not the same as fine), with battery percent and input volts. PET/CT and Nuc Med units show only rooms and power. The demo fleet has no such units, so these are not pictured.
Table view

Your choice of Cards or Table is remembered in that browser. With rooms or power in the fleet, a second table, Environment & power, appears under Magnet channels.
Messages you may see
- Reconnecting… showing the last readings received: the network dropped. The page recovers on its own.
- Your session has expired: press Sign in again.
- No assets match the current filter: click the active tile again.
- A red banner beginning Alerting has stopped, Alerting is hitting errors, Nobody receives alert email or Alert email is not going out (admins only): follow its link to Admin → Company → Alert email.
Asset page
Click any card, table row or ticker entry. The page refreshes every 30 seconds. Sections appear top to bottom in the order below.
1. Header

The badge reads Cryogenics nominal, Cryogenics: watch, Cryogenics: action needed or Cryogenics: critical. The lines under it say why, for example Coldhead 58.2 K; nominal is about 4 K.
2. Alerts

Engineers and admins see who owns each open alert, and the same Acknowledge / mute buttons as the Alerts page. See Acting on an alert.
3. Live readings

A water tile may show a dash with the note device reports 5.024 / fixed placeholder. The MagMon sends a fixed number when no sensor is fitted on that input. The app treats it as no reading and keeps it out of the trend.
4. Device error codes

CS1 is the compressor status. Zero is healthy. EC codes are the device's active errors. Codes 2, 6 and 26 are dimmed and tagged common across the fleet: they appear on many healthy units. The full lists are in Reference.
5. Outside conditions

This is context for the chilled-water loop: a hot afternoon shows in water temperature before it shows in helium. Weather never opens an alert.
6. Helium forecast

| Field | Meaning |
|---|---|
| Headline | Helium rising, Helium stable, Slow decline, ~N days/weeks/months to refill or Refill due now. |
| Rate | Percent of helium gained or lost per day. |
| Now (fit) | Today's level on the fitted line. Smoother than the live reading. |
| Days left | Days until the line reaches 30 %. |
| Fit R² | How straight the trend is, 0 to 1. Higher is more trustworthy. |
| Confidence | High needs 3 days of data and R² of 0.6. Medium needs 1 day and 0.3. Otherwise low: treat the date as a rough guide. |
7. Trends


The window is fixed at 24 hours. There is no range selector.
8. Recent readings

Units with rooms or a UPS also show zone cards (Section 1: Engineering, Section 2: Tech / Patient, Section 3: Equipment) and a Power card with the state, battery percent, input volts and whether mains power was lost in the last 24 hours.
Alerts

The three groups
- Needs someone: open, and nobody has picked it up. Work this list first.
- Acknowledged: still faulted, but someone owns it. It notifies again if it gets worse.
- Muted: conditions someone has silenced for a while. They still open alerts and still show on the fleet. They only stop email and push.
Within a group, critical comes before warning, and the oldest comes first. Acknowledging or muting never removes an alert. It leaves only when the condition clears. The page refreshes every 60 seconds.
Reading a row
| Part | Meaning |
|---|---|
| Asset name | Click to open the unit. |
| Kind | What sort of alert it is. See the table below. |
| critical | The higher of two severities. No tag means warning. |
| new | Opened since the last 9:00 AM cutoff. |
| muted | Not emailed or pushed. |
| open N days | Shown from one day. Turns amber at three days. |
| re-raised N× | Notified again because it stayed open and unacknowledged. |
| Left edge | Red for critical or no-data kinds, amber otherwise, gray when muted. |
Alert kinds
| Kind | What happened |
|---|---|
| Threshold | A reading crossed a rule set in Admin. |
| Out of range | A reading is outside its built-in healthy range. |
| Offline | No new reading for longer than the unit's offline limit. |
| Reporting stalled | The collector is reachable but no new data arrives. |
| Never reported | The unit was added but has never sent a reading. |
| MagMon silent / Env silent | One of the unit's two collectors has stopped while the other still reports. |
| Sensor fault | A water sensor is absent or has failed. |
| Cooling fault | Chilled-water flow has collapsed. |
| Helium loss | Helium is falling faster than normal boil-off. |
| Trending | A reading is heading for a limit, with days remaining. |
| Flatlined | A reading has been exactly constant, so the sensor is probably not measuring. |
| Unlike the fleet | A reading differs from the other units. Recorded only: never emailed or pushed. |
| Router offline / Pi offline | Network information. Never emailed or pushed. |
An alert opens on the third bad reading within five minutes and resolves after 15 clean minutes, so a single blip does not page anyone. A power outage opens on the first reading.
Acting on an alert
- Press Acknowledge / mute on the alert.
- Optionally type a note in Note (what did you find?), for example flow sensor loose at the pump, service booked Thursday.
- Choose one:
- Accept: it is real and you own it. It still pages if it gets worse.
- Mute…: it is real and you are living with it. Stop paging for a while.
- False alarm: the alert itself is wrong and the limit needs tuning. An admin sees the flag.
Muting
- Press Mute… and type a reason in Why?. It is required: it is what the next person reads.
- Pick a length: 24 hours, 3 days, 1 week, 2 weeks. Admins also get Until I clear it.
- A mute covers one kind of alert on one reading of one unit.
- It stops email and push. The unit still shows as faulted everywhere.
- A rise from warning to critical breaks the mute and pages anyway. An alert muted while already critical stays silent until the mute ends.
- When a mute ends or is lifted, the alert pages again if it is still true.
Changing your mind
Edit note revises the note. Un-acknowledge returns the alert to Needs someone. Un-mute lifts a mute, from the alert or from its card in the Muted group. A mute card shows the time left, the reason and who set it.
Acting from an alert email
Each alert email carries an Acknowledge or mute link. It opens a small page that needs no sign-in and works once.
- Acknowledge: I have it
- False alarm
- Mute for 1 day, 3 days or 1 week. The note becomes the mute reason.
Opening the link changes nothing until you press a button. If the alert has already cleared, or someone else used the link, the page says so.
Morning debrief

Before 9:00 AM the page still shows the previous window. It rolls over on its own.
The sections
- Still open: open at this morning's cutoff. These need someone. Older alerts are included.
- Cleared overnight: open when the window began, recovered during it.
- Came and went: opened and resolved inside the window. No action, but watch for repeats.
Repeats of one condition collapse into one row marked ×N. Tags: cleared since means it resolved after the cutoff, and muted means it never emailed or pushed. An empty day reads All quiet.
The 9 AM email
The same summary is emailed once each morning, shortly after 9:00 AM Central, to every address ticked for Morning debrief in Admin → Company → Alert email. The subject says either all quiet or how many alerts are still open. Asset names in the email link to their pages.
TV wall

Two ways to open it
Signed in: press TV in the menu.
A dedicated screen with no login (admin sets it up once):
- In Admin → Company → Wall displays, type where the screen is in Where is this screen? and press Create display link.
- Copy the link straight away with Copy link. It is shown once and not stored.
- Open the link in the TV's own browser.
- On the page Use this screen as a wall display? press Use this screen.
That browser now shows one company's fleet, read-only, for a year. Do this on the TV, not on your own computer. To stop a screen, press Revoke on its row in Wall displays.
The header
- Five tallies: Alarm Warning Nominal No data Maint.
- Speaker button: the alarm chime. Off by default, remembered per browser.
- Exit button: leaves TV mode and returns to the fleet.
- Full-screen button.
- Clock and date, with live (green) or reconnecting (amber).
The chime
Click the speaker once to turn it on. You hear a short blip. From then on a four-note chime plays whenever a unit newly goes into alarm. It stays quiet for units already in alarm when the page opened, for warnings, and while disconnected. Browsers block sound until that one click, so a TV needs it after each restart of its browser.
Rotation
- Units in alarm take the first cells and never rotate.
- The remaining cells page through everyone else: warning, nominal, no data, then maintenance. Dots at the bottom show the page.
- Each page stays 15 seconds. The grid sizes itself to the screen: up to 4 columns and 4 rows.
- Each card shows Coldhead, Helium, He Press and Shield with sparklines, its faults or All readings nominal, and any refill chip.

Tuning a screen from its address
There are no keyboard shortcuts. Add settings to the end of the address, for example /tv?dwell=20&handoff=30.
| Setting | Effect | Range |
|---|---|---|
dwell=N | Seconds per page | 5–120, default 15 |
perView=N | Columns | 1–6 |
rows=N | Rows | 1–4 |
anim=… | Page transition: zoom, fade, slide, up, blur, none | default zoom |
handoff=N | Show the Shift Summary every N minutes | 1–240, off by default |
handoffSecs=N | Seconds the summary stays up | 5–120, default 18 |
Shift Summary

handoff set, this card appears on a schedule: units in alarm and offline by name, the nominal count, the lowest helium and the next refill.If the wall goes gray
The grid dims and a banner explains. This display has lost its connection recovers on its own when the network returns. This display is signed out means the link was revoked or expired: pair the screen again.
Account and push notifications
Open Account. The top line shows your username and role.
Change your username
Type it in New username and press Save.
Change your PIN
- Type your Current PIN.
- Type the New PIN. It must meet the PIN rules.
- Press Update PIN. A wrong current PIN counts toward the sign-in lockout.
Get alerts as push notifications
Push works per device. Repeat these steps on each phone or computer that should be notified.
- iPhone or iPad only: open the app in Safari, press Share, then Add to Home Screen. Open it from the new icon.
- Open Account. Under Notifications press Enable push.
- Allow notifications when the browser asks.
- The line now reads Push on. Press Send test to check.
Turn off stops push on that device. Push blocked means notifications were denied: allow them for the site in the browser's settings. Every member with push enabled receives every alert for the company. There is no per-person filter.
Sign out
Press Sign out at the bottom of the page.
Admin
Four short videos cover this section. Users, Companies and Activity are described in words: those tabs show real names.
Open Admin. Six tabs: Assets, Alerts, Users, Company, Companies (superadmin only) and Activity. Everything applies to the active company. Results appear as a short Done or Error message at the bottom.
Assets: add a unit
- Press + Add asset.
- Step 1, the unit. Choose the Type of unit: MRI (has a MagMon), PET/CT or Nuc Med (rooms and power only). Type the Asset tag: letters, digits, hyphen and underscore, up to 32 characters. Add the site name and address. The address drives the weather panel.
- Press Next: connect the collector.
- Step 2, the collector. For an MRI, type the MagMon's address on the site network. Port, username and password are pre-filled with the device's factory values: change them only if the site did. Set the Pi login the collector runs as, and Offline after (minutes without a reading), default 30.
- Press Add asset and get install script. The install panel opens. Or press Add without connecting to finish later: the unit then shows as never reported.
Assets: each row
| Button | What it does |
|---|---|
| Maintenance / End maintenance | Flags the unit as in service: blue, no pills, rules not evaluated. Use for a unit that is deliberately warm or down. |
| Get install script | Opens the three downloads for this unit's collector. |
| Edit | Type, tag, site, address, offline limit, MagMon address and login, service user. After changing the MagMon details, download the unit's config again and reinstall it. |
| Delete | Removes the unit and all its history. Cannot be undone. To keep history, use Maintenance instead. |
Each row also shows the collector version: plain when current, amber behind, current is X when the Pi needs an update, or version unknown for a very old collector.
Site location. Under the address, the row shows where the site was placed for weather. Set location or Change lets you pin latitude and longitude by hand. Leave both blank to mark the site as having no weather. Look up again retries the automatic lookup.
Assets: the install panel
- Poll interval (min), Service user, and for MRI units a Collector toggle: MagMon or Environmental.
- Three downloads: the collector script (the same for every unit), this unit's config, and this unit's systemd unit.
- Rotate token issues a new secret for the unit. The Pi stops reporting until you download the new config and reinstall it. Collected data is not affected.
- Download all (N) on the toolbar builds one zip for the whole fleet.
The unit's secret token exists only inside the downloaded config. Treat that file as a password.
Alerts: rules
A rule says: for this scope, when this reading meets this condition, open an alert.
- Choose the Scope: All assets for a company-wide default, or one asset for an override on that unit.
- Choose the Metric: helium level, helium pressure, water flow, water temp, shield temp, CS1/compressor, a zone's temperature or humidity, On UPS battery, UPS battery remaining, or UPS input voltage.
- Choose the Condition (
<<=>>==!=) and type the Threshold. - Leave Enabled ticked and press Add rule.
- Rules are checked every minute. Units in maintenance are skipped.
- A per-asset rule replaces the company rule for that reading and direction only.
- Severity is not a setting. An alert is critical when the value is past the built-in critical line, and always for a stopped compressor or a power outage. Otherwise it is a warning.
- Room temperature and humidity have no default limits. Set them per site.
Recent alerts below the rules lists the last 100 events, read-only.
Users
- Add a person: username, PIN and role (Viewer, Engineer, Admin), then Add user.
- Per person: tick TV access and Docs, change the role, or Reset PIN. A reset signs the person out everywhere.
- The Docs runbook holds real network details. Grant it only to installers.
Superadmins see every person across all companies instead, with Rename, Deactivate, Delete and a role per company. Choosing No access removes the person from that company.
Company
Company profile
Company name, the product name shown at the top of every page and on the wall, the small heading above page titles, and the logo (PNG, JPEG or WebP, up to 2 MB). Press Save company details. The logo saves as soon as you pick the file.
Alert email: recipients
- Type the Email address. It does not need to belong to a user.
- Tick Alert emails, Morning debrief, or both.
- Press Add recipient, then Test on the new row and confirm the email arrives.
Row buttons turn alerts or the debrief on and off per address, or remove it. The SMS channel can be stored but does not send yet.
Alert email: sender identity
Sender name, Reply-To and subject prefix, with a live preview. The sending address itself is a superadmin setting.
Wall displays
Create, revoke and delete links for dedicated screens. See TV wall.
Companies
- + Add company: name, URL slug (fixed once created), product name, heading and tagline.
- Edit: add the logo and an invite code. Anyone with the code can register into the company. Leave it empty to close self-registration.
- Disable: members lose access, wall screens go dark and alerts stop. Readings keep being collected.
- Delete: type the company name to confirm. Destroys its units, history, rules and display links.
Activity
The last 200 audit entries: sign-ins, failed attempts, and every change to units, people and rules, with who did it. Press Refresh to update.
Handbook and Docs
Handbook (admins) shows the fleet's state right now, how a reading travels from device to phone, what runs on a schedule, and what each alert means. Docs is the installer runbook: building a Pi, joining the VPN, wiring sensors and a UPS, site networking and troubleshooting. Both live inside the app and stay current with it.
Bringing a unit online
The short version. The full procedure, with the networking cases, is in the in-app Docs. Replace <ASSET> with the asset tag and <user> with the service user.
- Prepare the Pi. Raspberry Pi OS Lite 64-bit, hostname
<asset>-pi, SSH on. Update it, then join it to the VPN. - Add the unit in Admin → Assets and download its three files.
- Copy the files to the Pi.
scp nm-magmon-gateway.py <ASSET>-magmon.json nm-magmon-gateway-<ASSET>.service <user>@<host>:~/
- Install the script (once per Pi).
sudo apt-get install -y python3-requests sudo install -o root -g root -m 755 nm-magmon-gateway.py /opt/nm-magmon-gateway.py - Install the unit's config and service.
sudo install -d -m 755 /etc/nm-collector sudo install -o <user> -g "$(id -gn <user>)" -m 600 <ASSET>-magmon.json /etc/nm-collector/<ASSET>-magmon.json sudo install -m 644 nm-magmon-gateway-<ASSET>.service /etc/systemd/system/nm-magmon-gateway-<ASSET>.service sudo systemctl daemon-reload && sudo systemctl enable --now nm-magmon-gateway-<ASSET> - Check it.
systemctl status nm-magmon-gateway-<ASSET> --no-pager journalctl -u nm-magmon-gateway-<ASSET> -n 30 --no-pagerThe log should say starting for asset and then reported N sample(s). The card turns Online within a couple of minutes.
crontab -l on the Pi should not mention the collector.Room sensors and a UPS
The environmental collector reads up to three temperature and humidity sensors over RS-485 (addresses 1, 2, 3 for Engineering, Tech / Patient, Equipment) and a UPS over USB. On an MRI unit, open Get install script and set Collector to Environmental. PET/CT and Nuc Med units get it automatically. The install mirrors the steps above with nm-env-gateway.py, <ASSET>-env.json and nm-env-gateway-<ASSET>.service. An MRI unit can run both collectors side by side.
Later changes
- New collector version: replace
/opt/nm-magmon-gateway.pyand restart the services. One file updates every unit on that Pi. - New MagMon address, poll interval or rotated token: download the config again, reinstall it with mode 600, restart that unit's service.
Reference
What a healthy magnet looks like
| Reading | Typical | Alarm line |
|---|---|---|
| Helium level | 55–82 % | below 30 % |
| Coldhead | 3.9–4.3 K | above 20 K |
| Helium pressure | about 1.0 psi | above 4.5 psi |
| Water flow | about 2 gpm | set by rule (fleet default below 0.6) |
| Shield | varies by magnet | above 80 K |
| Compressor (CS1) | 0 | anything above 0 |
How fresh is what I am looking at?
| Thing | Refreshes |
|---|---|
| Collector reading | Every minute |
| Fleet, asset page, TV wall | Every 30 seconds while visible |
| Alerts page | Every 60 seconds |
| Debrief page | Every 5 minutes |
| Refill forecast, weather | Every 5 to 10 minutes |
| Alert rules checked | Every minute |
| Minute-by-minute history kept | 7 days. Daily summaries are kept for 400 days. |
| Unanswered alert re-raised | After 3 days, then every 7 days |
Compressor status (CS1)
| Code | Meaning |
|---|---|
| 0 | Running normally |
| 4, 14 | Klixon (thermal cut-out) error |
| 8, 18 | Compressor fuse tripped |
| 10 | Stopped, reason unknown |
| 11 | Stopped due to overheat |
| 12 | Stopped due to low helium pressure |
| 13 | Overheat and low pressure |
| 20, 30–33 | Powered off or cable disconnected (no 24 V signal). With a cold coldhead this reads as signal lost, not stopped. |
Common device error codes (EC)
| Code | Device text |
|---|---|
| 2 | He level too low (common across the fleet) |
| 6 | Water flow for compressor 1 too high (common across the fleet) |
| 23 | Coldhead RuO temp too hot |
| 26 | Vessel pressure too low (common across the fleet) |
| 121 | Compressor 1 is not running |
Groups: 5–8 water, 23–24 coldhead, 44–52 cable disconnects, 121–126 compressor. The asset page always shows the device's own text next to each code.
Glossary
- MagMon
- The magnet's own monitoring controller. The collector reads it.
- Collector
- The program on a Raspberry Pi at the site that reads the equipment and reports once a minute.
- Coldhead
- The cryocooler stage that keeps helium liquid. About 4 K when healthy.
- Shield
- The magnet's thermal shield. Warms early when cooling is lost.
- CS1
- Compressor status code. Zero is healthy.
- iR305
- The cellular router at some sites.
- Tailscale
- The private network that links each Pi to the service.
- UPS
- Battery backup. On battery means mains power is lost.
When something looks wrong
| You see | Likely cause | Do this |
|---|---|---|
| Offline, with IR305 Down | The site's cellular router or its signal | Contact the site about the router's power and antenna. |
| Offline, with Tailscale Down | The Pi is off or has no network | Have the site check the Pi's power and cable. |
| Reporting stalled | The Pi is up, but the collector is not delivering new readings | Installer: check the service log on the Pi. See Docs, Run and troubleshoot. |
| MagMon silent | The Pi cannot reach the MagMon | Check the cable between the Pi and the MagMon, and the MagMon's address in Admin. |
| Never reported | The collector was never installed, or its config is wrong | Finish the install, or download the config again. |
| Compressor stopped and coldhead rising | A real cooling loss | Treat as urgent. Check chilled water and compressor power on site. |
| Compressor signal lost, coldhead still about 4 K | The status cable, not the compressor | Schedule a cable check. Cooling is still working. |
| Water … sensor blind | A water sensor is absent or failed | Mute with a reason if known and accepted, and book the repair. |
| A reading flatlined | A stuck sensor | Do not trust that reading. Book a sensor check. |
| No alert emails | No recipients, or the sender is not working | Admin → Company → Alert email. Add the address and press Test. |
| No push on a phone | Push is per device, and iPhones need the Home Screen app | Account → Enable push on that device, then Send test. |
| Page says Not updating | Your own connection | It recovers on its own. Reload if it persists. |
| Locked out | Five wrong PINs | Wait 15 minutes, or ask an admin to reset the PIN. |