User guide

Help

Every screen, control, color and number in the app, in the order you meet them. Operators can stop after Account. Admins and installers read on. Screenshots and videos come from the demo fleet (MM-1001 to MM-1012). Click any screenshot to enlarge it.

Start here

Magnet Monitor watches MRI magnets and the rooms around them. A small collector at each site reads the magnet's monitor (the MagMon), and optionally room sensors and a UPS, once a minute. The app shows the fleet, opens alerts when a reading goes wrong, and emails or pushes them to whoever is on call.

The five-minute tour

  1. Open the Fleet page. Units that need attention are sorted to the top. Red is an alarm, amber a warning, green nominal, gray no data, blue maintenance.
  2. Read the pills on a card. Each pill names one fault and its value, for example Helium critical 22.9 %.
  3. Click the card. The asset page shows the open alerts, live readings, device error codes, the helium refill forecast and 24 hours of trends.
  4. Open Alerts. Anything under Needs someone has no owner. Engineers and admins acknowledge or mute from here.
  5. Check the Debrief each morning. It lists what opened, cleared and stayed open since 9:00 AM yesterday.
Chapter 1, 1:52. The Fleet page: tiles, colors, pills, chips and the table view.

Who can do what

Each person has one role per company. Two extra grants, TV and Docs, are ticked per person by an admin.

Screen or actionViewerEngineerAdmin
Fleet, asset pages, Alerts, Debrief, AccountYesYesYes
Acknowledge, mute, flag false alarmNoYes (mute up to 2 weeks)Yes (any length)
TV wallWith TV grantWith TV grantYes
Docs (installer runbook)With Docs grantWith Docs grantYes
Admin panel and HandbookNoNoYes

A superadmin is an admin in every company, and also sees the Companies tab and the company switcher.

Signing in

Everyone
Sign-in screen with Username, PIN and Keep me signed in
Sign in. Username, PIN, and the stay-signed-in tick box.
Create account screen with Invite code field
Create account. Adds the Invite code field.

Sign in with an existing account

  1. Type your Username and PIN.
  2. Leave Keep me signed in on this device ticked on your own device: you stay signed in for 30 days. Untick it on a shared computer: the session ends when the browser closes, or after one day.
  3. Press Sign in.

Create an account with an invite code

  1. Press Need an account? Register with an invite code.
  2. Choose a username and a PIN, and type the invite code your admin gave you.
  3. Press Create account. You are signed in as a Viewer. An admin can raise your role afterward.

PIN rules

  • At least 6 characters, at most 72. Letters are allowed.
  • Not one character repeated (111111).
  • Not a simple run up or down (123456, 987654, abcdef).
  • Not a common choice such as 123123, password or qwerty.

An older, weaker PIN still signs you in, but every page shows a banner asking you to change it under Account.

Wrong PIN and lockout

Any failed sign-in shows the same message: Invalid username or PIN. After several wrong attempts, sign-in from this network pauses for 15 minutes. Five wrong PINs within 15 minutes pause that account from your network for 15 minutes. Wait, or ask an admin to reset the PIN.

View live demo at the top right opens the simulated fleet without signing in.

Fleet dashboard

Everyone
Fleet dashboard: header, four status tiles, alert ticker, view toggle and unit cards
The dashboard. Status line, four filter tiles, the alert ticker, the Cards/Table toggle, then one card per unit.

The status line

Under the title: a green dot and Live, the time of the last update, and the count of assets and sites. The page refreshes itself every 30 seconds while the tab is visible. If it reads Not updating with an amber dot, the last successful refresh was over 75 seconds ago and the readings on screen may be old.

The four tiles

All assets, Online, Offline, Unknown tiles
Click a tile to show only those units. Click it again to show all.

The tiles count connectivity only, not faults.

  • Online: a new reading arrived within the unit's offline limit (30 minutes unless an admin changed it).
  • Offline: the last new reading is older than that limit.
  • Unknown: the unit has never sent a reading.
Dashboard filtered to offline units
The Offline tile selected: only the four offline units remain.

The alert ticker

The red strip scrolls every unit in alarm or warning with its faults. Hover to pause it. Click a name to open that unit. It disappears when the fleet is clean.

Order of the cards

Alarm first, then warning, nominal, no data, and maintenance last. Within a group the order is alphabetical. There is no manual sort or search.

Reading a card

MM-1002 card in alarm with two red pills
Alarm. Red rail and ring. Pills: Compressor stopped code 33, Coldhead hot 58 K. Footer: Online, Cryo Critical.
MM-1008 card nominal with 3w to refill chip and Cryo Watch
Nominal with a refill chip. Green rail. Blue chip: 3w to refill. Footer: Cryo Watch.
MM-1007 card in maintenance, blue
Maintenance. Blue rail. Readings are abnormal but no pills show, because the unit is flagged as in service.
MM-1009 card unknown, never reported
No data. Gray rail. Never reported: the collector has not been connected yet.
Part of the cardWhat it tells you
Name and siteThe asset tag and where it is. The small weather icon and temperature are the outside conditions at the site (US National Weather Service). Information only.
Colored rail and ringThe alarm level: Alarm Warning Nominal No data Maintenance. The ring's dot pulses while the unit is online.
PillsUp to three faults, worst first. The asset page shows all of them.
Six tilesHelium level (%), Water flow (gpm), Helium pressure (psi), Water temp (°F), Shield temp (K), Compressor (Running, Stopped or Signal lost, with its code). Each has a one-hour sparkline. A dash means no reading.
Footer chipsConnectivity (Online, Offline, Unknown), any collector or network chip, and the Cryo chip when the cryogenics are not nominal.
Bottom rightAge of the last new reading, or never reported.

Fault pills and their limits

These are the built-in limits. They are checked only while the unit is online. An admin's alert rules can move the warning lines for a company or a single unit.

PillShown whenLevel
Compressor stopped code NCS1 above 0Alarm
Compressor signal lost code NCS1 is 20, 30, 31, 32 or 33 and coldhead still at or below 6 KWarning
Coldhead hotabove 20 KAlarm
Coldhead highabove 6 KWarning
Helium criticalbelow 30 %Alarm
Helium lowbelow 50 %Warning
Helium highabove 90 %Warning
He pressure highabove 4.5 psiAlarm
He pressure abnormalbelow 0 or above 3 psiWarning
Shield hotabove 80 KAlarm
Shield warmabove 60 KWarning
Offline ageNo new reading for longer than the offline limitAlarm
Reporting stalled ageThe collector's Pi is reachable, but no new data is arrivingAlarm

Other pills come from open alerts: sensor blind (no reading from a water sensor), Cooling fault, trending (days to a limit), Helium falling (% per day), flatlined (a stuck value), a channel alarm from a rule, and POWER OUTAGE on UPS battery, which is always an alarm.

Footer chips

MM-1011 offline card with IR305 Down chip
IR305 Down. The site's cellular router is offline, so the whole site has lost its link.
MM-1012 offline card with Tailscale Down chip
Tailscale Down. The collector's Pi has dropped off the VPN: the Pi or its own connection is down.
ChipMeaning
Online / Offline / UnknownConnectivity, as counted by the tiles.
MagMon silentThe unit is online, but magnet readings have stopped arriving.
MagMon not connectedThe unit has never sent magnet readings.
Env silentRoom sensor and UPS readings have stopped.
iR305 / iR305 downState of the site's cellular router. Shown only where the app can see the router.
Tailscale / Tailscale downState of the collector Pi's VPN link.
Cryo watch / urgent / critical / no dataThe cryogenics grade. No chip means nominal.

The router and VPN chips explain why a unit is offline. They never open an alert of their own.

The Cryo grade

One judgment of the cold system, for MRI units only.

ReadingWatchUrgentCritical
Helium levelbelow 60 %below 50 %below 30 %
Coldhead5.0 K or more6.0 K or more20 K or more
Vessel pressure2.0 psi or moreabove 3.0 psiabove 4.5 psi
Coldhead minus recondenser-2.0 K or more-
Two of coldhead, recondenser, shield at 100 K or more--Magnet is warm

The refill chip

The droplet chip estimates when helium will reach 30 %. It uses a straight-line fit of the last 7 days, restarted after any refill. It appears only when helium is clearly falling.

  • refill due: under one day.
  • 5d to refill: under 14 days. The chip is amber.
  • 3w to refill, 3mo to refill: further out. The chip is blue.
  • No chip: stable, rising, or 120 days or more away.

Rooms and power (units with sensors or a UPS)

Units fitted with room sensors add one column per zone, Engineering, Tech / Patient and Equipment, each with temperature (°F) and humidity (%RH). Units with a UPS add a power row: Wall power (green), ON BATTERY (red) or Power unknown (gray: the UPS is not answering, which is not the same as fine), with battery percent and input volts. PET/CT and Nuc Med units show only rooms and power. The demo fleet has no such units, so these are not pictured.

Table view

Fleet in table view
One row per unit. The colored edge and dot give the level, the line under the name gives the reason and age, and the offending cell is tinted red or amber. Click a row to open the unit.

Your choice of Cards or Table is remembered in that browser. With rooms or power in the fleet, a second table, Environment & power, appears under Magnet channels.

Messages you may see

  • Reconnecting… showing the last readings received: the network dropped. The page recovers on its own.
  • Your session has expired: press Sign in again.
  • No assets match the current filter: click the active tile again.
  • A red banner beginning Alerting has stopped, Alerting is hitting errors, Nobody receives alert email or Alert email is not going out (admins only): follow its link to Admin → Company → Alert email.

Asset page

EveryoneTriage: engineer, admin

Click any card, table row or ticker entry. The page refreshes every 30 seconds. Sections appear top to bottom in the order below.

Chapter 2, 1:25. Opening a unit: the verdict, alerts, readings, codes, forecast and trends.

1. Header

Asset header for MM-1002 with pills and cryogenics badge
Site, asset tag, address and weather, connectivity, every fault pill, then the Cryogenics badge with its reasons in plain words.

The badge reads Cryogenics nominal, Cryogenics: watch, Cryogenics: action needed or Cryogenics: critical. The lines under it say why, for example Coldhead 58.2 K; nominal is about 4 K.

2. Alerts

Alerts section with two active and two recently resolved
Open alerts first, with how long each has been open. Recently resolved shows the last five, with how long each lasted.

Engineers and admins see who owns each open alert, and the same Acknowledge / mute buttons as the Alerts page. See Acting on an alert.

3. Live readings

Six reading tiles
The same six readings as the card, without sparklines.

A water tile may show a dash with the note device reports 5.024 / fixed placeholder. The MagMon sends a fixed number when no sensor is fitted on that input. The app treats it as no reading and keeps it out of the trend.

4. Device error codes

Device error codes: CS1 33
Codes reported by the MagMon itself, with the device's own wording.

CS1 is the compressor status. Zero is healthy. EC codes are the device's active errors. Codes 2, 6 and 26 are dimmed and tagged common across the fleet: they appear on many healthy units. The full lists are in Reference.

5. Outside conditions

Outside conditions panel
Temperature, humidity, wind and observation time from the nearest weather station. Needs a site address.

This is context for the chilled-water loop: a hot afternoon shows in water temperature before it shows in helium. Weather never opens an alert.

6. Helium forecast

Helium forecast: about 3 weeks to refill
MM-1008: about 3 weeks to refill, losing 1.32 % a day, high confidence.
FieldMeaning
HeadlineHelium rising, Helium stable, Slow decline, ~N days/weeks/months to refill or Refill due now.
RatePercent of helium gained or lost per day.
Now (fit)Today's level on the fitted line. Smoother than the live reading.
Days leftDays until the line reaches 30 %.
Fit R²How straight the trend is, 0 to 1. Higher is more trustworthy.
ConfidenceHigh needs 3 days of data and R² of 0.6. Medium needs 1 day and 0.3. Otherwise low: treat the date as a rough guide.

7. Trends

Six trend charts for the last 24 hours
One chart per reading: the last 24 hours as 15-minute averages. The dashed line on Water temp is outside air, on the right-hand scale.
Trend chart with hover tooltip
Hover, or touch on a phone, for the time, the value and how many samples it averages.

The window is fixed at 24 hours. There is no range selector.

8. Recent readings

Recent readings table
The same 15-minute averages as a table, newest first. Samples is how many one-minute readings went into each row: about 15 is healthy.

Units with rooms or a UPS also show zone cards (Section 1: Engineering, Section 2: Tech / Patient, Section 3: Equipment) and a Power card with the state, battery percent, input volts and whether mains power was lost in the last 24 hours.

Alerts

Read: everyoneTriage: engineer, admin
Alerts page with four tallies and Needs someone list
Alerts. Four tallies, then every open alert grouped by who owns it.
Chapter 3, 1:37. Alerts: who owns what, and Accept, Mute and False alarm.

The three groups

  • Needs someone: open, and nobody has picked it up. Work this list first.
  • Acknowledged: still faulted, but someone owns it. It notifies again if it gets worse.
  • Muted: conditions someone has silenced for a while. They still open alerts and still show on the fleet. They only stop email and push.

Within a group, critical comes before warning, and the oldest comes first. Acknowledging or muting never removes an alert. It leaves only when the condition clears. The page refreshes every 60 seconds.

Reading a row

PartMeaning
Asset nameClick to open the unit.
KindWhat sort of alert it is. See the table below.
criticalThe higher of two severities. No tag means warning.
newOpened since the last 9:00 AM cutoff.
mutedNot emailed or pushed.
open N daysShown from one day. Turns amber at three days.
re-raised N×Notified again because it stayed open and unacknowledged.
Left edgeRed for critical or no-data kinds, amber otherwise, gray when muted.

Alert kinds

KindWhat happened
ThresholdA reading crossed a rule set in Admin.
Out of rangeA reading is outside its built-in healthy range.
OfflineNo new reading for longer than the unit's offline limit.
Reporting stalledThe collector is reachable but no new data arrives.
Never reportedThe unit was added but has never sent a reading.
MagMon silent / Env silentOne of the unit's two collectors has stopped while the other still reports.
Sensor faultA water sensor is absent or has failed.
Cooling faultChilled-water flow has collapsed.
Helium lossHelium is falling faster than normal boil-off.
TrendingA reading is heading for a limit, with days remaining.
FlatlinedA reading has been exactly constant, so the sensor is probably not measuring.
Unlike the fleetA reading differs from the other units. Recorded only: never emailed or pushed.
Router offline / Pi offlineNetwork information. Never emailed or pushed.

An alert opens on the third bad reading within five minutes and resolves after 15 clean minutes, so a single blip does not page anyone. A power outage opens on the first reading.

Acting on an alert

EngineerAdmin
  1. Press Acknowledge / mute on the alert.
  2. Optionally type a note in Note (what did you find?), for example flow sensor loose at the pump, service booked Thursday.
  3. Choose one:
    • Accept: it is real and you own it. It still pages if it gets worse.
    • Mute…: it is real and you are living with it. Stop paging for a while.
    • False alarm: the alert itself is wrong and the limit needs tuning. An admin sees the flag.

Muting

  1. Press Mute… and type a reason in Why?. It is required: it is what the next person reads.
  2. Pick a length: 24 hours, 3 days, 1 week, 2 weeks. Admins also get Until I clear it.
  • A mute covers one kind of alert on one reading of one unit.
  • It stops email and push. The unit still shows as faulted everywhere.
  • A rise from warning to critical breaks the mute and pages anyway. An alert muted while already critical stays silent until the mute ends.
  • When a mute ends or is lifted, the alert pages again if it is still true.

Changing your mind

Edit note revises the note. Un-acknowledge returns the alert to Needs someone. Un-mute lifts a mute, from the alert or from its card in the Muted group. A mute card shows the time left, the reason and who set it.

Mute or Maintenance? Mute silences one condition's notifications and leaves the unit red. Maintenance is an admin switch on the whole unit: it turns the unit blue, removes its pills and stops rules being evaluated for it. Use Maintenance for a unit that is deliberately warm or out of service.

Acting from an alert email

Each alert email carries an Acknowledge or mute link. It opens a small page that needs no sign-in and works once.

  • Acknowledge: I have it
  • False alarm
  • Mute for 1 day, 3 days or 1 week. The note becomes the mute reason.

Opening the link changes nothing until you press a button. If the alert has already cleared, or someone else used the link, the page says so.

Morning debrief

Everyone
Morning debrief with four tallies and Still open list
Morning debrief. The window under the title is always 9:00 AM yesterday to 9:00 AM today, Central time.
Chapter 4, 0:44. The morning debrief.

Before 9:00 AM the page still shows the previous window. It rolls over on its own.

The sections

  • Still open: open at this morning's cutoff. These need someone. Older alerts are included.
  • Cleared overnight: open when the window began, recovered during it.
  • Came and went: opened and resolved inside the window. No action, but watch for repeats.

Repeats of one condition collapse into one row marked ×N. Tags: cleared since means it resolved after the cutoff, and muted means it never emailed or pushed. An empty day reads All quiet.

The 9 AM email

The same summary is emailed once each morning, shortly after 9:00 AM Central, to every address ticked for Morning debrief in Admin → Company → Alert email. The subject says either all quiet or how many alerts are still open. Asset names in the email link to their pages.

TV wall

AdminTV grantPaired screens
TV wall: header tallies, alert ticker and six alarm cards
The wall. Built to be read from across a room. Units in alarm are pinned to the first cells and flash.
Chapter 5, 0:50. The TV wall: pinned alarms, rotation and the chime.

Two ways to open it

Signed in: press TV in the menu.

A dedicated screen with no login (admin sets it up once):

  1. In Admin → Company → Wall displays, type where the screen is in Where is this screen? and press Create display link.
  2. Copy the link straight away with Copy link. It is shown once and not stored.
  3. Open the link in the TV's own browser.
  4. On the page Use this screen as a wall display? press Use this screen.

That browser now shows one company's fleet, read-only, for a year. Do this on the TV, not on your own computer. To stop a screen, press Revoke on its row in Wall displays.

The header

  • Five tallies: Alarm Warning Nominal No data Maint.
  • Speaker button: the alarm chime. Off by default, remembered per browser.
  • Exit button: leaves TV mode and returns to the fleet.
  • Full-screen button.
  • Clock and date, with live (green) or reconnecting (amber).

The chime

Click the speaker once to turn it on. You hear a short blip. From then on a four-note chime plays whenever a unit newly goes into alarm. It stays quiet for units already in alarm when the page opened, for warnings, and while disconnected. Browsers block sound until that one click, so a TV needs it after each restart of its browser.

Rotation

  • Units in alarm take the first cells and never rotate.
  • The remaining cells page through everyone else: warning, nominal, no data, then maintenance. Dots at the bottom show the page.
  • Each page stays 15 seconds. The grid sizes itself to the screen: up to 4 columns and 4 rows.
  • Each card shows Coldhead, Helium, He Press and Shield with sparklines, its faults or All readings nominal, and any refill chip.
TV wall second page
A later page of the rotation.

Tuning a screen from its address

There are no keyboard shortcuts. Add settings to the end of the address, for example /tv?dwell=20&handoff=30.

SettingEffectRange
dwell=NSeconds per page5–120, default 15
perView=NColumns1–6
rows=NRows1–4
anim=…Page transition: zoom, fade, slide, up, blur, nonedefault zoom
handoff=NShow the Shift Summary every N minutes1–240, off by default
handoffSecs=NSeconds the summary stays up5–120, default 18

Shift Summary

Shift Summary overlay
With handoff set, this card appears on a schedule: units in alarm and offline by name, the nominal count, the lowest helium and the next refill.

If the wall goes gray

The grid dims and a banner explains. This display has lost its connection recovers on its own when the network returns. This display is signed out means the link was revoked or expired: pair the screen again.

Account and push notifications

Everyone
Chapter 6, 0:55. Help, push notifications and the daily routine.

Open Account. The top line shows your username and role.

Change your username

Type it in New username and press Save.

Change your PIN

  1. Type your Current PIN.
  2. Type the New PIN. It must meet the PIN rules.
  3. Press Update PIN. A wrong current PIN counts toward the sign-in lockout.

Get alerts as push notifications

Push works per device. Repeat these steps on each phone or computer that should be notified.

  1. iPhone or iPad only: open the app in Safari, press Share, then Add to Home Screen. Open it from the new icon.
  2. Open Account. Under Notifications press Enable push.
  3. Allow notifications when the browser asks.
  4. The line now reads Push on. Press Send test to check.

Turn off stops push on that device. Push blocked means notifications were denied: allow them for the site in the browser's settings. Every member with push enabled receives every alert for the company. There is no per-person filter.

Sign out

Press Sign out at the bottom of the page.

Admin

Admin

Four short videos cover this section. Users, Companies and Activity are described in words: those tabs show real names.

Admin 1, 1:10. Units: adding one, and the four buttons on every row.
Admin 2, 0:45. Alert rules.
Admin 3, 0:52. Company, alert email and wall displays.
Admin 4, 0:34. People and the audit log.

Open Admin. Six tabs: Assets, Alerts, Users, Company, Companies (superadmin only) and Activity. Everything applies to the active company. Results appear as a short Done or Error message at the bottom.

Assets: add a unit

  1. Press + Add asset.
  2. Step 1, the unit. Choose the Type of unit: MRI (has a MagMon), PET/CT or Nuc Med (rooms and power only). Type the Asset tag: letters, digits, hyphen and underscore, up to 32 characters. Add the site name and address. The address drives the weather panel.
  3. Press Next: connect the collector.
  4. Step 2, the collector. For an MRI, type the MagMon's address on the site network. Port, username and password are pre-filled with the device's factory values: change them only if the site did. Set the Pi login the collector runs as, and Offline after (minutes without a reading), default 30.
  5. Press Add asset and get install script. The install panel opens. Or press Add without connecting to finish later: the unit then shows as never reported.

Assets: each row

ButtonWhat it does
Maintenance / End maintenanceFlags the unit as in service: blue, no pills, rules not evaluated. Use for a unit that is deliberately warm or down.
Get install scriptOpens the three downloads for this unit's collector.
EditType, tag, site, address, offline limit, MagMon address and login, service user. After changing the MagMon details, download the unit's config again and reinstall it.
DeleteRemoves the unit and all its history. Cannot be undone. To keep history, use Maintenance instead.

Each row also shows the collector version: plain when current, amber behind, current is X when the Pi needs an update, or version unknown for a very old collector.

Site location. Under the address, the row shows where the site was placed for weather. Set location or Change lets you pin latitude and longitude by hand. Leave both blank to mark the site as having no weather. Look up again retries the automatic lookup.

Assets: the install panel

  • Poll interval (min), Service user, and for MRI units a Collector toggle: MagMon or Environmental.
  • Three downloads: the collector script (the same for every unit), this unit's config, and this unit's systemd unit.
  • Rotate token issues a new secret for the unit. The Pi stops reporting until you download the new config and reinstall it. Collected data is not affected.
  • Download all (N) on the toolbar builds one zip for the whole fleet.

The unit's secret token exists only inside the downloaded config. Treat that file as a password.

Alerts: rules

A rule says: for this scope, when this reading meets this condition, open an alert.

  1. Choose the Scope: All assets for a company-wide default, or one asset for an override on that unit.
  2. Choose the Metric: helium level, helium pressure, water flow, water temp, shield temp, CS1/compressor, a zone's temperature or humidity, On UPS battery, UPS battery remaining, or UPS input voltage.
  3. Choose the Condition (< <= > >= = !=) and type the Threshold.
  4. Leave Enabled ticked and press Add rule.
  • Rules are checked every minute. Units in maintenance are skipped.
  • A per-asset rule replaces the company rule for that reading and direction only.
  • Severity is not a setting. An alert is critical when the value is past the built-in critical line, and always for a stopped compressor or a power outage. Otherwise it is a warning.
  • Room temperature and humidity have no default limits. Set them per site.

Recent alerts below the rules lists the last 100 events, read-only.

Users

  • Add a person: username, PIN and role (Viewer, Engineer, Admin), then Add user.
  • Per person: tick TV access and Docs, change the role, or Reset PIN. A reset signs the person out everywhere.
  • The Docs runbook holds real network details. Grant it only to installers.

Superadmins see every person across all companies instead, with Rename, Deactivate, Delete and a role per company. Choosing No access removes the person from that company.

Company

Company profile

Company name, the product name shown at the top of every page and on the wall, the small heading above page titles, and the logo (PNG, JPEG or WebP, up to 2 MB). Press Save company details. The logo saves as soon as you pick the file.

Alert email: recipients

  1. Type the Email address. It does not need to belong to a user.
  2. Tick Alert emails, Morning debrief, or both.
  3. Press Add recipient, then Test on the new row and confirm the email arrives.

Row buttons turn alerts or the debrief on and off per address, or remove it. The SMS channel can be stored but does not send yet.

Alert email: sender identity

Sender name, Reply-To and subject prefix, with a live preview. The sending address itself is a superadmin setting.

Wall displays

Create, revoke and delete links for dedicated screens. See TV wall.

Companies

Superadmin
  • + Add company: name, URL slug (fixed once created), product name, heading and tagline.
  • Edit: add the logo and an invite code. Anyone with the code can register into the company. Leave it empty to close self-registration.
  • Disable: members lose access, wall screens go dark and alerts stop. Readings keep being collected.
  • Delete: type the company name to confirm. Destroys its units, history, rules and display links.

Activity

The last 200 audit entries: sign-ins, failed attempts, and every change to units, people and rules, with who did it. Press Refresh to update.

Handbook and Docs

Handbook (admins) shows the fleet's state right now, how a reading travels from device to phone, what runs on a schedule, and what each alert means. Docs is the installer runbook: building a Pi, joining the VPN, wiring sensors and a UPS, site networking and troubleshooting. Both live inside the app and stay current with it.

Bringing a unit online

AdminInstaller

The short version. The full procedure, with the networking cases, is in the in-app Docs. Replace <ASSET> with the asset tag and <user> with the service user.

  1. Prepare the Pi. Raspberry Pi OS Lite 64-bit, hostname <asset>-pi, SSH on. Update it, then join it to the VPN.
  2. Add the unit in Admin → Assets and download its three files.
  3. Copy the files to the Pi.
    scp nm-magmon-gateway.py <ASSET>-magmon.json nm-magmon-gateway-<ASSET>.service <user>@<host>:~/
  4. Install the script (once per Pi).
    sudo apt-get install -y python3-requests
              sudo install -o root -g root -m 755 nm-magmon-gateway.py /opt/nm-magmon-gateway.py
  5. Install the unit's config and service.
    sudo install -d -m 755 /etc/nm-collector
              sudo install -o <user> -g "$(id -gn <user>)" -m 600 <ASSET>-magmon.json /etc/nm-collector/<ASSET>-magmon.json
              sudo install -m 644 nm-magmon-gateway-<ASSET>.service /etc/systemd/system/nm-magmon-gateway-<ASSET>.service
              sudo systemctl daemon-reload && sudo systemctl enable --now nm-magmon-gateway-<ASSET>
  6. Check it.
    systemctl status nm-magmon-gateway-<ASSET> --no-pager
              journalctl -u nm-magmon-gateway-<ASSET> -n 30 --no-pager
    The log should say starting for asset and then reported N sample(s). The card turns Online within a couple of minutes.
Run it under systemd, never cronTwo copies of a collector fight over the same unit. crontab -l on the Pi should not mention the collector.

Room sensors and a UPS

The environmental collector reads up to three temperature and humidity sensors over RS-485 (addresses 1, 2, 3 for Engineering, Tech / Patient, Equipment) and a UPS over USB. On an MRI unit, open Get install script and set Collector to Environmental. PET/CT and Nuc Med units get it automatically. The install mirrors the steps above with nm-env-gateway.py, <ASSET>-env.json and nm-env-gateway-<ASSET>.service. An MRI unit can run both collectors side by side.

Later changes

  • New collector version: replace /opt/nm-magmon-gateway.py and restart the services. One file updates every unit on that Pi.
  • New MagMon address, poll interval or rotated token: download the config again, reinstall it with mode 600, restart that unit's service.

Reference

What a healthy magnet looks like

ReadingTypicalAlarm line
Helium level55–82 %below 30 %
Coldhead3.9–4.3 Kabove 20 K
Helium pressureabout 1.0 psiabove 4.5 psi
Water flowabout 2 gpmset by rule (fleet default below 0.6)
Shieldvaries by magnetabove 80 K
Compressor (CS1)0anything above 0

How fresh is what I am looking at?

ThingRefreshes
Collector readingEvery minute
Fleet, asset page, TV wallEvery 30 seconds while visible
Alerts pageEvery 60 seconds
Debrief pageEvery 5 minutes
Refill forecast, weatherEvery 5 to 10 minutes
Alert rules checkedEvery minute
Minute-by-minute history kept7 days. Daily summaries are kept for 400 days.
Unanswered alert re-raisedAfter 3 days, then every 7 days

Compressor status (CS1)

CodeMeaning
0Running normally
4, 14Klixon (thermal cut-out) error
8, 18Compressor fuse tripped
10Stopped, reason unknown
11Stopped due to overheat
12Stopped due to low helium pressure
13Overheat and low pressure
20, 30–33Powered off or cable disconnected (no 24 V signal). With a cold coldhead this reads as signal lost, not stopped.

Common device error codes (EC)

CodeDevice text
2He level too low (common across the fleet)
6Water flow for compressor 1 too high (common across the fleet)
23Coldhead RuO temp too hot
26Vessel pressure too low (common across the fleet)
121Compressor 1 is not running

Groups: 5–8 water, 23–24 coldhead, 44–52 cable disconnects, 121–126 compressor. The asset page always shows the device's own text next to each code.

Glossary

MagMon
The magnet's own monitoring controller. The collector reads it.
Collector
The program on a Raspberry Pi at the site that reads the equipment and reports once a minute.
Coldhead
The cryocooler stage that keeps helium liquid. About 4 K when healthy.
Shield
The magnet's thermal shield. Warms early when cooling is lost.
CS1
Compressor status code. Zero is healthy.
iR305
The cellular router at some sites.
Tailscale
The private network that links each Pi to the service.
UPS
Battery backup. On battery means mains power is lost.

When something looks wrong

You seeLikely causeDo this
Offline, with IR305 DownThe site's cellular router or its signalContact the site about the router's power and antenna.
Offline, with Tailscale DownThe Pi is off or has no networkHave the site check the Pi's power and cable.
Reporting stalledThe Pi is up, but the collector is not delivering new readingsInstaller: check the service log on the Pi. See Docs, Run and troubleshoot.
MagMon silentThe Pi cannot reach the MagMonCheck the cable between the Pi and the MagMon, and the MagMon's address in Admin.
Never reportedThe collector was never installed, or its config is wrongFinish the install, or download the config again.
Compressor stopped and coldhead risingA real cooling lossTreat as urgent. Check chilled water and compressor power on site.
Compressor signal lost, coldhead still about 4 KThe status cable, not the compressorSchedule a cable check. Cooling is still working.
Water … sensor blindA water sensor is absent or failedMute with a reason if known and accepted, and book the repair.
A reading flatlinedA stuck sensorDo not trust that reading. Book a sensor check.
No alert emailsNo recipients, or the sender is not workingAdmin → Company → Alert email. Add the address and press Test.
No push on a phonePush is per device, and iPhones need the Home Screen appAccount → Enable push on that device, then Send test.
Page says Not updatingYour own connectionIt recovers on its own. Reload if it persists.
Locked outFive wrong PINsWait 15 minutes, or ask an admin to reset the PIN.