Live Demo · Remote Monitoring

Alerts

Everything open across your units, and who is handling it.

11needs someone
0acknowledged
4critical
0muted

What happened overnight is in the morning debrief.

Needs someone 11

Open and unowned. Nobody has picked these up.

MM-1005Thresholdcritical

MM-1005 Helium 16.32 % — below 50 %

since 8/17/2026, 7:06:00 PM · open 97 days

MM-1002Thresholdcritical

MM-1002 compressor stopped — code 33: compressor powered off or cable disconnected (no 24v signal). The coldhead is at 58.23 K, which is consistent with a compressor that really has stopped

since 8/17/2026, 7:06:00 PM · open 97 days

MM-1002Out of rangecritical

MM-1002 Coldhead at 57.86 K — nominal is below 6 K

since 8/20/2026, 3:40:44 AM · open 94 days

MM-1008Helium losscritical

MM-1008 helium is falling 1.273 %/day from 36.9 % — its own baseline over the previous fortnight was 0.000 %/day, and a healthy magnet on this fleet moves under 0.1 % in three weeks. At this rate it reaches 30 % in about 5 days. The coldhead is at 4.30 K, so the cryocooler is running — this reads as venting or a leak rather than boil-off

since 11/13/2026, 12:00:00 AM · open 9 days

MM-1006Offline

MM-1006 offline — no report since 2026-08-17 15:31 UTC

since 8/17/2026, 7:06:00 PM · open 97 days

MM-1012Offline

MM-1012 offline — no report since 2026-08-17 16:53 UTC

since 8/17/2026, 7:06:00 PM · open 97 days

MM-1011Offline

MM-1011 offline — no report since 2026-08-17 17:37 UTC

since 8/17/2026, 7:06:00 PM · open 97 days

MM-1004Offline

MM-1004 offline — no report since 2026-08-17 18:18 UTC

since 8/17/2026, 7:06:00 PM · open 97 days

MM-1009Never reported

MM-1009 has never reported telemetry since it was added on 2026-08-17 — check the collector install on its Pi

since 8/19/2026, 10:08:32 PM · open 95 days

MM-1005Helium loss

MM-1005 helium is falling 0.608 %/day from 15.2 % — its own baseline over the previous fortnight was 0.000 %/day, and a healthy magnet on this fleet moves under 0.1 % in three weeks. The coldhead is at 4.43 K, so the cryocooler is running — this reads as venting or a leak rather than boil-off

since 11/13/2026, 12:00:00 AM · open 9 days

MM-1008Threshold

MM-1008 Helium 49.88 % — below 50 %

since 11/13/2026, 3:31:00 AM · open 9 days