Live Demo · Remote Monitoring
Morning debrief
Sat, Nov 21, 9:00 AM → Sun, Nov 22, 9:00 AM CST
Still open11
Open as of this morning's cutoff. These need someone.
MM-1008 Helium 49.88 % — below 50 %
MM-1005 helium is falling 0.608 %/day from 15.2 % — its own baseline over the previous fortnight was 0.000 %/day, and a healthy magnet on this fleet moves under 0.1 % in three weeks. The coldhead is at 4.43 K, so the cryocooler is running — this reads as venting or a leak rather than boil-off
MM-1008 helium is falling 1.273 %/day from 36.9 % — its own baseline over the previous fortnight was 0.000 %/day, and a healthy magnet on this fleet moves under 0.1 % in three weeks. At this rate it reaches 30 % in about 5 days. The coldhead is at 4.30 K, so the cryocooler is running — this reads as venting or a leak rather than boil-off
MM-1002 Coldhead at 57.86 K — nominal is below 6 K
MM-1009 has never reported telemetry since it was added on 2026-08-17 — check the collector install on its Pi
MM-1006 offline — no report since 2026-08-17 15:31 UTC
MM-1012 offline — no report since 2026-08-17 16:53 UTC
MM-1011 offline — no report since 2026-08-17 17:37 UTC
MM-1004 offline — no report since 2026-08-17 18:18 UTC
MM-1005 Helium 16.32 % — below 50 %
MM-1002 compressor stopped — code 33: compressor powered off or cable disconnected (no 24v signal). The coldhead is at 58.23 K, which is consistent with a compressor that really has stopped
Covers every alert the monitor opened or resolved in the window, plus anything still open from before it. Rolls over automatically at 9:00 AM CST.